FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2050595 · Received March 15, 2011

Report

Report Number
2531779-2011-01601
Event Type
Malfunction
Date Received
March 15, 2011
Date of Event
February 14, 2011
Report Date
February 16, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE, THE RESULTS WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS. DURING EVALUATION THE PUMP BOOTED TO AN AUDIBLE ALARM AND AN EPROM FAILURE OCCURRED. THE EPROM FAILURE STILL PERSISTED AFTER THE INVESTIGATOR ATTEMPTED TO UPLOAD SOFTWARE FILES. THE PUMP COULD NOT BE DOWNLOADED DUE TO A SOFTWARE CORRUPTION.

Description of Event or Problem · 1

THE DISTRIBUTOR REPORTED THAT THE PUMP DID NOT START.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1