FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 2050428
·
Received April 5, 2011
Report
- Report Number
- 2531779-2011-02264
- Event Type
- Malfunction
- Date Received
- April 5, 2011
- Report Date
- March 8, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Removal / Correction Number
- 2531779-03/24/2010-003-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. DURING TESTING, PUMP WAS LOADED AND PRIMED SUCCESSFULLY. DURING EVALUATION, THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION.
Description of Event or Problem · 1
PUMP IS RETURNED FOR A SHORT LOAD STEP CAUSING A LARGE PRIME VOLUME. EVALUATION REVEALED OUT OF CALIBRATION FORCE SENSOR. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |