FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 2050425
·
Received April 5, 2011
Report
- Report Number
- 2531779-2011-02281
- Event Type
- Malfunction
- Date Received
- April 5, 2011
- Date of Event
- March 8, 2011
- Report Date
- March 8, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION CONFIRMED THAT THE UP ARROW AND DOWN ARROW KEYPAD BUTTONS ARE INTERMITTENTLY RESPONSIVE. INVESTIGATION REVEALED THAT THE UP ARROW AND DOWN ARROW BUTTON KEY CONTACTS WERE MISALIGNED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THE PATIENT REPORTED THAT THE UP ARROW AND DOWN ARROW BUTTONS BECAME INTERMITTENTLY UNRESPONSIVE THIS EVENING, AND REQUIRE SEVERAL PRESSES TO OBTAIN A RESPONSE. HE NOTED THAT THE BUTTONS FEEL HARD WHEN PRESSED. THE PATIENT DENIED PHYSICAL DAMAGE TO THE RUBBER KEYPAD; DENIED EXPOSING THE PUMP TO WATER; AND STATED THAT HE WEARS THE PUMP IN HIS POCKET.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONE TOUCH PING |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |