FDA Adverse Event Injury Summary report: N

MINIMED QUICK-SET

MDR report key: 20503212 · Received October 22, 2024

Report

Report Number
3003442380-2024-28704
Event Type
Injury
Date Received
October 22, 2024
Date of Event
September 21, 2024
Report Date
July 29, 2025
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244017573
PMA / PMN Number
K160648
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

E1: PATIENT CITY: (B)(6). PATIENT COUNTRY: UNITED STATES.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS - H11: INVESTIGATION SUMMARY: COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED FOR RECORD (B)(4) ON 28-JUL-2025. COMPLAINT INVESTIGATION: THE REFERENCE SAMPLES CANNOT BE TESTED BECAUSE THERE WAS NO SPECIFIC MALFUNCTION TO INVESTIGATE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6002729 WAS MANUFACTURED ACCORDING TO THE DOCUMENT VERSION 77 ON THE PACKING PROCESS IN THE MACHINE 12, ON 14-AUG-2023, WITH A TOTAL OF (B)(4) UNITS. TRENDING: A QUERY WAS RUN IN DATABASE ON 28-JUL-2025 AGAINST HARM CODE SIGNS OF UNTREATED HYPOGLYCEMIA THAT PATIENT IS UNABLE TO SELF-MANAGE REQUIRING INTERVENTION BY A HEALTH CARE PROFESSIONAL (HCP) OR REQUIRES EMERGENCY ADVANCED LIFE, MALFUNCTION CODE NO MALFUNCTION DESCRIBES AND LOT 6002729 AND NO MORE COMPLAINT HAS BEEN REGISTERED IN DATABASE FOR THE SAME LOT, HARM CODE AND MALFUNCTION CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION, ONLY ONE COMPLAINT RECEIVED ON THE LOT IN QUESTION AND HARM CODE, THE TEST ON REFERENCE SAMPLES WERE NOT TESTED SINCE A NO MALFUNCTION RELATED TO THE INFUSION SET WAS REPORTED, THEREFORE NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED AN EVENT OF HYPOGLYCEMIA BECAUSE PATIENT GAVE HIMSELF TOO MUCH INSULIN USING MANUAL INJECTION ON (B)(6)2024 WHICH LEADS TO HOSPITALIZATION. NO FURTHER INFORMATION WAS AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1946671 MINIMED QUICK-SET UNO QUICK-SET 60/9 SC1 MECA FPA UNOMEDICAL A/S MMT-397A 6002729 05705244017573

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention| H