FDA Adverse Event Injury Summary report: N

CONTAK RENEWAL

MDR report key: 2050315 · Received April 12, 2011

Report

Report Number
2124215-2011-04653
Event Type
Injury
Date Received
April 12, 2011
Date of Event
February 16, 2011
Report Date
February 16, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT IN OUR LABORATORY, DEVICE INTERROGATION REVEALED THAT THE DEVICE HAD REACHED ELECTIVE REPLACEMENT INDICATOR (ERI). A LONGEVITY CALCULATION USING THE MOST RECENT PROGRAMMED PARAMETERS AND THERAPY USE INFORMATION INDICATED THAT THE DEVICE FELL SHORT OF LABELED LONGEVITY EXPECTATIONS. REVIEW OF SELF-DIAGNOSTIC TEST RESULTS STORED IN DEVICE MEMORY DID NOT IDENTIFY ANY PERIODS OF EXCESSIVE CURRENT OR ENERGY USE DURING THE LIFE OF THE DEVICE. PRELIMINARY TESTING OF DEVICE FUNCTION DEMONSTRATED NORMAL BEHAVIOR. FURTHER ANALYSIS OF DEVICE MEMORY INDICATED THAT AN ERI INDICATOR WAS DECLARED DUE TO TWO CONSECUTIVE CHARGES EXCEEDING THE ERI CHARGE TIME LIMIT. THE MONITORING VOLTAGE WAS ABOVE THE ERI LIMIT, INDICATING THAT THE BATTERY WAS NOT DEPLETED AND HAD SUFFICIENT CAPACITY REMAINING TO PROVIDE THERAPY IF NEEDED. HOWEVER, THE CHARGE TIMES WOULD HAVE BEEN LONGER THAN THE 18 SECONDS ERI CHARGE TIME LIMIT SPECIFIED IN DEVICE LABELING. DEVICE CIRCUITRY WAS DESIGNED SUCH THAT THE ERI CHARGE TIME LIMIT OF 18 SECONDS WOULD BE REACHED NEAR THE CORRESPONDING ERI MONITORING VOLTAGE LIMIT. IN THIS CASE, A NORMAL BUT EARLIER THAN EXPECTED BUILDUP OF INTERNAL BATTERY IMPEDANCE INCREASED CHARGE TIME, CAUSING THE CHARGE TIME INDICATOR TO TRIP ERI EARLY. DESPITE THE LONGEVITY SHORTFALL, REPLACEMENT INDICATORS OPERATED AS DESIGNED AND THE DEVICE WAS CAPABLE OF DETECTING AND TREATING ARRHYTHMIAS FOR SEVERAL ADDITIONAL MONTHS AS DESCRIBED IN DEVICE LABELING. THIS ENSURED PATIENT SAFETY DURING THE REPLACEMENT PERIOD.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS EXPLANTED DUE TO BATTERY STATUS AT ELECTIVE REPLACEMENT INDICATOR (ERI). THERE WAS A CONCERN THAT THE BATTERY HAD DEPLETED EARLIER THAN EXPECTED. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H155

Patients

Seq Age Sex Outcome Treatment
1