ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2011-05769
- Event Type
- Injury
- Date Received
- April 12, 2011
- Date of Event
- February 18, 2011
- Report Date
- February 18, 2011
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- PMA / PMN Number
- P910073
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
TO DATE, INFORMATION SUGGESTS THE CRT-D AND LV LEAD REMAIN ACTIVELY IN SERVICE, BUT THIS RV LEAD WAS SURGICALLY ABANDONED AS A RESULT OF LEAD FRACTURE. IF NEW INFORMATION WOULD BECOME AVAILABLE, THIS REPORT WILL BE FURTHER EVALUATED AND UPDATED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) IN ASSOCIATION WITH THIS TRANSVENOUS RIGHT VENTRICULAR (RV) LEAD AND THE LEFT VENTRICULAR (LV) LEAD DID EXHIBIT PACING IMPEDANCE GREATER THAN 2000 OHMS ON BOTH THE RV AND LV LEADS. THIS WAS EVIDENT TO HAVE INCREASED OVER TIME. THE LV LEAD HAD BEEN PROGRAMMED TIP TO COIL AND WHEN THE COIL WAS REMOVED FROM THE SYSTEM CONFIGURATION, IMPEDANCE WAS NORMAL. RV THRESHOLD HAD RISEN TO 5 VOLTS AT 1 MILLISECOND. SHOCK LEAD IMPEDANCE WAS EVIDENT TO HAVE INCREASED ABOUT 50% FROM THE 60S TO THE RANGE OF 90 OHMS PACE IMPEDANCE. A STORED EVENT WAS IDENTIFIED WHICH INDICATED RV NOISE, CAUSING OVERSENSING. THIS NOISE WAS UNABLE TO BE REPRODUCED WITH ISOMETRICS. NO INAPPROPRIATE THERAPY HAD BEEN DELIVERED, BUT THE PATIENT DID REPORT MORE FATIGUE AND SHORTNESS OF BREATH. IT WAS NOT KNOWN AT THE TIME OF THIS INITIAL REPORT WHETHER ACTUAL ASYSTOLE GREATER THAN 2 BEATS/SECONDS HAD OCCURRED. A CONDUCTOR FRACTURE OF THE RV LEAD WAS SUSPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | NVY | CPI - DEL CARIBE | 0138 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | 4542| 0138| 4480| H227| 4136 |