FDA Adverse Event Injury Summary report: N

VITALITY

MDR report key: 2050248 · Received April 12, 2011

Report

Report Number
2124215-2011-03525
Event Type
Injury
Date Received
April 12, 2011
Date of Event
February 16, 2011
Report Date
January 8, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE AVAILABLE INFORMATION SUGGESTS THIS DEVICE REMAINS IN SERVICE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

THIS DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED INDICATING A CHARGE TIME OUT WAS OBSERVED. THE DEVICE WAS EXPLANTED AND SUCCESSFULLY REPLACED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DECLARED END OF LIFE APPROXIMATELY 50 MONTHS AFTER IMPLANT. IT WAS REPORTED THAT THE PATIENT WAS LOST TO FOLLOW-UP FOR APPROXIMATELY 28 MONTHS. THE MOST RECENT MONITORING VOLTAGE WAS 2.59 V AND THE CHARGE TIME WAS 35.1 SECONDS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T135

Patients

Seq Age Sex Outcome Treatment
1 65 YR Hospitalization| L| R T135| 0158