VITALITY
Report
- Report Number
- 2124215-2011-03525
- Event Type
- Injury
- Date Received
- April 12, 2011
- Date of Event
- February 16, 2011
- Report Date
- January 8, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE AVAILABLE INFORMATION SUGGESTS THIS DEVICE REMAINS IN SERVICE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS REPORT WILL BE UPDATED.
THIS DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS REPORT WILL BE UPDATED.
ADDITIONAL INFORMATION WAS RECEIVED INDICATING A CHARGE TIME OUT WAS OBSERVED. THE DEVICE WAS EXPLANTED AND SUCCESSFULLY REPLACED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DECLARED END OF LIFE APPROXIMATELY 50 MONTHS AFTER IMPLANT. IT WAS REPORTED THAT THE PATIENT WAS LOST TO FOLLOW-UP FOR APPROXIMATELY 28 MONTHS. THE MOST RECENT MONITORING VOLTAGE WAS 2.59 V AND THE CHARGE TIME WAS 35.1 SECONDS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T135 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Hospitalization| L| R | T135| 0158 |