FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2050243 · Received April 12, 2011

Report

Report Number
1423500-2011-04368
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
March 23, 2011
Report Date
March 23, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING INVESTIGATION BY BAXTER THIS INCIDENT WAS DETERMINED TO BE CAUSED BY USE/USER ERROR AND THERE WAS NO ALLEGATION OF A PRODUCT MALFUNCTION. THEREFORE A BATCH REVIEW AND SAMPLE EVALUATION WILL NOT BE CONDUCTED. A FOLLOW-UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION IS OBTAINED.

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT IS FOR A REPORT OF A USER ERROR. THE PATIENT CHANGED OUT THE CASSETTE AND REUSED THE SOLUTION BAGS. THIS COMPLAINT WAS NOT CONFIRMED. PER THE COMPLAINT INFORMATION, THE CAUSE OF THE COMPLAINT IS USE ERROR. THE LABEL REVIEW FOUND THE LABELING ADEQUATE FOR THE POTENTIAL USE ERROR IN THIS COMPLAINT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM .THE ROOT CAUSE INVESTIGATION IS IN PROGRESS.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER TO REPORT A CHECK LINES AND BAGS ALARM WHICH OCCURRED ON HOME CHOICE (HC) DURING USE DURING PRIME. THE HP STATED THAT HE TRIED TO TROUBLESHOOT HC, BUT COULD NOT CLEAR THE ALARM. THE HP STATED THAT HE POWER CYCLED THE HC AND REPLACED CASSETTE WITH A NEW ONE BUT DID NOT REPLACE THE BAGS. THE BAXTER TECHNICAL SERVICE REPRESENTATIVE (TSR) INSTRUCTED THE HP TO START OVER WITH ALL NEW SUPPLIES. THE HC WAS OPERATIONAL. THERE WAS PATIENT INVOLVEMENT. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 62 YR HOMECHOICE