FDA Adverse Event Malfunction Summary report: N

VITALITY

MDR report key: 2050185 · Received April 12, 2011

Report

Report Number
2124215-2011-02936
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
February 16, 2011
Report Date
March 1, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST-MARKET QUALITY ASSURANCE LABORATORY, A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW-VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION.

Additional Manufacturer Narrative · 1

THIS PRODUCT IS CURRENTLY UNDERGOING DETAILED ANALYSIS. THIS EVENT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED DUE TO NORMAL BATTERY DEPLETION AND WAS RETURNED FOR DISPOSAL. ROUTINE INITIAL DEVICE TESTING INDICATED THAT DETAILED ANALYSIS WAS REQUIRED. TO DATE, THERE HAVE BEEN NO REPORTED ADVERSE PATIENT EFFECTS RELATED TO THIS CLINICAL OBSERVATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T135

Patients

Seq Age Sex Outcome Treatment
1 82 YR E102| T135| 0158