FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 2050182 · Received April 1, 2011

Report

Report Number
1644487-2011-00691
Event Type
Malfunction
Date Received
April 1, 2011
Date of Event
March 2, 2011
Report Date
March 2, 2011
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

IT WAS INITIALLY REPORTED THAT THE "CFCARD" FOLDER WAS NOT ABLE TO BE SEEN ON THE FLASHCARD THAT WAS BEING USED TO UPGRADE THE HANDHELD. THE COMPANY REP PERFORMED A HARD RESET BUT THE ISSUE REMAINED. ANOTHER FLASHCARD WAS USED TO TRY TO UPGRADE THE HANDHELD AND WAS SUCCESSFUL. THE FLASHCARD WAS RETURNED TO THE MFR FOR EVAL. PRODUCT ANALYSIS IS PLANNED BUT HAS NOT BEEN COMPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING SOFTWARE LYJ CYBERONICS, INC. 250 834042

Patients

Seq Age Sex Outcome Treatment
1