FDA Adverse Event Malfunction Summary report: N

AFFINITY IV BED

MDR report key: 2050163 · Received April 5, 2011

Report

Report Number
1824206-2011-02013
Event Type
Malfunction
Date Received
April 5, 2011
Date of Event
March 11, 2011
Report Date
March 11, 2011
Manufacturer
HILL-ROM, INC.
Product Code
HDD
Removal / Correction Number
Z-0548-2010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE BRAKE DETENT WAS SHATTERED. THE TECHNICIAN PERFORMED THE MOD AND ADJUSTED ALL BRAKE CASTERS TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

ACCOUNT ALLEGED THAT THE BED WAS STUCK IN STEER. NO REPORT OF INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY IV BED BIRTHING BED HDD HILL-ROM, INC. 3700

Patients

Seq Age Sex Outcome Treatment
1