FDA Adverse Event Malfunction Summary report: N

AFFINITY IV BED

MDR report key: 2050161 · Received April 5, 2011

Report

Report Number
1824206-2011-02012
Event Type
Malfunction
Date Received
April 5, 2011
Date of Event
March 11, 2011
Report Date
March 11, 2011
Manufacturer
HILL-ROM, INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN PERFORMED THE MOD TO REPAIR THE BRAKE DETENT AND ADJUSTED ALL BRAKE CASTERS TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

ACCOUNT ALLEGED THAT THE BRAKE DETENT HAS FAILED. THE BRAKES WOULD LOCK BUT CASTERS WOULD STILL ROLL. NO REPORT OF INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY IV BED BIRTHING BED HDD HILL-ROM, INC. 3700

Patients

Seq Age Sex Outcome Treatment
1