OMNIPOD INSULIN PUMP
Report
- Report Number
- 3004464228-2011-00135
- Event Type
- Malfunction
- Date Received
- April 5, 2011
- Date of Event
- March 6, 2011
- Report Date
- March 6, 2011
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042792
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE INVESTIGATION OF THE RETURNED POD FOUND DISCOLORATION INSIDE THE DEVICE. RESIDUAL FLUID AND CORROSION WAS SEEN ON THE SURFACES OF INTERNAL ASSEMBLIES, WHICH IS EVIDENCE OF AN INTERNAL FLUID LEAK. A LEAK TEST WAS PERFORMED, WHICH CONFIRMED THE PRESENCE OF A LEAK IN THE FLUID PATH CAUSED BY A TEAR IN THE CANNULA TUBING. THE LEAK WOULD HAVE RESULTED IN INSULIN COMING INTO CONTACT WITH INTERNAL COMPONENTS AND ASSEMBLIES, ULTIMATELY CAUSING THE DEVICE TO FAIL. A POD MALFUNCTION, THEREFORE, IS CONFIRMED TO HAVE BEEN A CONTRIBUTING FACTOR TO THE CUSTOMER'S HIGH BG LEVELS. A REVIEW OF LOT QUALIFICATION RECORDS REVEALED EVIDENCE OF THIS FAILURE MODE. ULTIMATELY, LOT QUALIFICATION RESULTS WERE DEEMED ACCEPTABLE AND THE LOT PASSED THE ACCEPTANCE CRITERIA. INSULET HAS INITIATED AN INTERNAL INVESTIGATION INTO THIS PARTICULAR FAILURE MODE AND HAS TAKEN MULTIPLE ACTIONS TO MINIMIZE AND PREVENT ITS RECURRENCE. A RISK ASSESSMENT HAS ALSO BEEN CONDUCTED AS WELL AS ONGOING MONITORING TO ENSURE THAT THIS LEVEL OF FAILURE DOES NOT EXCEED ACTION LIMITS (AS DEFINED BY INSULET'S INTERNAL PROCEDURES). THE OCCURRENCE RATE OF THIS FAILURE MODE IS LESS THAN 4/100,000; IMPROVEMENT ACTIVITIES ARE IN-PROCESS. LABELING, TRAINING AND CLINICAL ADVICE CONTINUES TO INDICATE THAT ONGOING BG MONITORING IS NECESSARY. ALTHOUGH WE DO NOT RELY ON LABELING, THIS IS ADD'L MITIGATION AS PART OF THE NORMAL ACTIVITIES OF A DIABETIC.
A REVIEW OF THE CUSTOMER'S BG HISTORY INDICATED THAT HIGH LEVELS (254-303MG/DL) WERE EXPERIENCED WITHIN THE FIRST DAY OF WEARING THE POD. MULTIPLE BOLUSES WERE ADMINISTERED DURING THIS TIME, BUT HER LEVELS REMAINED CONSISTENTLY HIGH. NO SPECIFIC DEVICE ISSUE WAS CITED. THE POD WILL BE RETURNED FOR EVAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 11200 | L30438 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR | Other |