FDA Adverse Event
Malfunction
Summary report: N
STEALTHSTATION TREON TREATMENT GUIDANCE SYSTEM
MDR report key: 2050138
·
Received April 1, 2011
Report
- Report Number
- 1723170-2011-00607
- Event Type
- Malfunction
- Date Received
- April 1, 2011
- Date of Event
- October 6, 2010
- Report Date
- October 6, 2010
- Manufacturer
- MEDTRONIC NAVIGATION, INC.
- Product Code
- HAW
- PMA / PMN Number
- K050438
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
PATIENT DEMOGRAPHIC UNKNOWN. PER RMA, NUMEROUS PARTS WERE SENT TO SITE TO RESOLVE ISSUE. ON RETURN OF PARTS WERE SENT TO SITE TO RESOLVE ISSUE. ON RETURN OF PARTS THE POWER SUPPLY SEEMED TO PROVIDE AN UNSTEADY 9 TO 10 VOLTS AND SPARKED DURING TESTING. POWER SUPPLY ALSO HAS BURNED PINS AT CONNECTOR. THE TREON MONITOR POWER PIGTAIL CONNECTOR IS SEVERELY BURNED AROUND RED WIRES ON FEMALE CONNECTORS. ALL OTHER PARTS WERE RETURNED UNUSED.
Description of Event or Problem · 1
SITE REPORTED THEIR MONITOR WENT BLACK ON THE TREON SYSTEM AFTER NATIONAL DISPLAY SYSTEM CAME UP. THE SURGEON DECIDED TO DISCONTINUE THE USE OF THE SYSTEM AND THE PROCEDURE WAS SUCCESSFUL WITHOUT IMPACT ON THE PATIENT OUTCOME.