BD SYRINGE 5ML LL BNS
Report
- Report Number
- 1213809-2024-00747
- Event Type
- Malfunction
- Date Received
- October 21, 2024
- Date of Event
- October 4, 2024
- Report Date
- January 21, 2025
- Manufacturer
- BECTON DICKINSON MEDICAL SYSTEMS
- Product Code
- FMF
- UDI-DI
- 50382903010276
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ONE PHOTO WAS PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THE PHOTO SHOWS ONE LOOSE SYRINGE WITH THE SCALE MARKINGS LIGHT AND MISSING ON THE BARREL. THE CONDITION OBSERVED IS NON-CONFORMING PER PRODUCT SPECIFICATION. POTENTIAL ROOT CAUSE FOR THE SCALE MARKINGS MISSING DEFECT IS ASSOCIATED WITH THE MARKING PROCESS. THESE CONDITIONS ARE OCCURRING BELOW THEIR EXPECTED FREQUENCY SO, NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 4152592. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. WE APPRECIATE YOU TAKING THE TIME TO BRING THIS OBSERVATION TO OUR ATTENTION. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.
NO ADDITIONAL INFORMATION WAS PROVIDED.
MATERIAL#: 301027, BATCH#: 4152592. IT WAS REPORTED THAT THE BD SYRINGE 5ML LL BNS HAS A SCALE MARKING ISSUE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: RCC RECEIVED A COMPLAINT VIA EMAIL. OUR PRODUCTION DEPARTMENT HAS DETECTED A SYRINGE WHERE THE PRINT IS MISSING OR ILLEGIBLE. THIS INFORMATION IS BEING SENT TO YOU AS AN AWARENESS EFFORT ONLY. UNFORTUNATELY, THE SYRINGE WAS DISCARDED AT OUR FACILITY AND IS UNAVAILABLE TO SEND TO YOU TO REVIEW. WE WERE ABLE TO OBTAIN A PHOTO OF THE SYRINGE BEFORE IT WAS DISCARDED (SEE BELOW). PLEASE FEE FREE TO CONTACT ME IF YOU HAVE ANY QUESTIONS. CPN 14858 ¿ YOUR PART#: 301027. YOUR LOT#: 4152592. PO: (B)(4). ADDITIONAL INFORMATION PROVIDED: 1. ANY ADVERSE EVENT OR SERIOUS INJURY REPORTED TO PATIENT OR HEALTHCARE PROFESSIONAL? IF YES, PLEASE PROVIDE THE DETAILS. NO. 2. CAN YOU PLEASE PROVIDE AN EXACT DATE OF EVENT IN FORMAT DD-MMM-YYYY? 04OCT2024. 3. TOTAL NUMBER OF OCCURRENCES? 4.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1300895 | BD SYRINGE 5ML LL BNS | PISTON SYRINGE | FMF | BECTON DICKINSON MEDICAL SYSTEMS | 4152592 | 50382903010276 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |