FDA Adverse Event Malfunction Summary report: N

SYRINGE 50ML LL TIP 1ML

MDR report key: 20494816 · Received October 21, 2024

Report

Report Number
1911916-2024-00745
Event Type
Malfunction
Date Received
October 21, 2024
Date of Event
October 16, 2024
Report Date
November 8, 2024
Manufacturer
BECTON DICKINSON
Product Code
FMF
UDI-DI
30382903096535
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PR (B)(4). FOLLOW UP FOR DEVICE EVALUATION. IT WAS REPORTED THERE IS BLURRY PRINTING ON THE SYRINGES. TO AID IN THE INVESTIGATION, SEVEN SAMPLES IN SEALED PACKAGING BLISTERS AND ONE PHOTO WAS PROVIDED FOR EVALUATION BY OUR QUALITY TEAM. A VISUAL INSPECTION WAS PERFORMED. THE SAMPLES HAVE A LIGHT SCALE MARKING PRINTING. THE PHOTO PROVIDED SHOWS A SYRINGE WITH NO PACKAGING BLISTER AND A LIGHT SCALE MARKING PRINT. THE LIGHT PRINTING IS CONSIDERED AN ACCEPTABLE IMPERFECTION. NO OTHER DEFECTS OR IMPERFECTIONS WERE OBSERVED. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 309653, LOT 4080153. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. THERE WERE NO RELATED QUALITY NOTIFICATIONS. ALL PROCESSES AND FINAL INSPECTIONS COMPLIED WITH SPECIFICATION REQUIREMENTS. BASED ON THE INVESTIGATION AND WITH THE RETURNED SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED, BUT ACCEPTABLE.

Additional Manufacturer Narrative · 0

INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED.

Description of Event or Problem · 0

THE LUER LOCK SYRINGE HAS BLURRY PRINTING ON THE MARKINGS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2220626 SYRINGE 50ML LL TIP 1ML SYRINGE, PISTON FMF BECTON DICKINSON 4080153 30382903096535

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown