CAPSURE Z
Report
- Report Number
- 2649622-2011-05212
- Event Type
- Injury
- Date Received
- April 11, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P850089/S33
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE ACTUAL LEAD WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. FIFTEEN - VENTRICULAR NST (NON-SUSTAINED TACHYCARDIA) <=210 MS AVERAGE V-CYCLE ON (B)(6) 2011 IN THE TIMEFRAME BETWEEN 07:19:47 AND 15:01:57. FOUR - VF (VENTRICULAR FIBRILLATION) <=200 MS BETWEEN (B)(6) 2011 08:19:39 AND (B)(6) 2011 17:41:45. VENTRICULAR SHORT INTERVAL COUNT V-SIC=14.9 COUNTS AVG/DAY, IN 7.99 DAYS, BETWEEN (B)(6) 2011 18:01:33 AND (B)(6) 2011 17:52:35. ONE - PATIENT ALERT FOR LEAD FAILURE PREDICTOR ON (B)(6) 2011 19:44:45.
IT WAS REPORTED THAT THE PATIENT WAS SEEN FOR SYNCOPE. EVALUATION ON THE LEAD SHOWED MANY NON-SUSTAINED TACHYCARDIA EPISODES AND ONE VENTRICULAR FIBRILLATION EPISODE SHOWING NOISE LEADING TO PACING INHIBITION AND ONE SHOCK. THE LEAD INTEGRITY ALERT TRIGGERED FOR SHORT INTERVAL COUNTS AND NON-SUSTAINED TACHYCARDIA EPISODES. THE LEAD WAS CAPPED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSURE Z | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5034 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | Hospitalization| L| R | 4194 IMPLANTABLE PACING LEAD| D224TRK IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 6949 IMPLANTABLE TACHY LEAD| 5534 IMPLANTABLE PACING LEAD |