FDA Adverse Event Malfunction Summary report: N

LIVONGO

MDR report key: 20489031 · Received October 19, 2024

Report

Report Number
3011420269-2024-00057
Event Type
Malfunction
Date Received
October 19, 2024
Date of Event
July 27, 2023
Report Date
August 29, 2023
Manufacturer
GUANGDONG TRANSTEK MEDICAL ELECTRONICS CO. LTD
Product Code
DXN
PMA / PMN Number
K131395
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

CAUSE: 1. TRANSTEK COULDN'T GET THE COMPLAINT DETAILS FROM END USER AND COULDN'T GET BACK THE DEVICE FOR FURTHER ANALYSIS. 2. NO INVENTORY OF THE SAME MODEL IS FOUND. 3. WE CHECKED THE ALL RELATED DHR, INCLUDING MANUFACTURED PROCESS RECORDS, FQC INSPECTION RECORDS, ECT, AND NO ACCURACY ISSUE WAS FOUND. 4. THE PRODUCT HAS PASSED CLINICAL VALIDATION ISO 81060-2:2018. ITS PERFORMANCE MEETS THE REQUIREMENTS. 5. THE INSTRUCTIONS HAVE ALREADY COVERED THE RELEVANT OPERATION. THE CUSTOMER DIDN'T READ THE INSTRUCTIONS CAREFULLY. CORRECTIVE ACTION: 1. ESTABLISH A REGULAR COMMUNICATION MECHANISM WITH CUSTOMERS. 2. PREPARING A DOCUMENT CONCERNING TROUBLESHOOTING AND ESSENTIAL PRECAUTIONS AS A NOTIFICATION OF REMINDER FOR CUSTOMER PROMOTION TO MINIMIZE THE RISK OF USER MISOPERATIONS. CONCLUSION: THIS ISSUE WOULD NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY, NOT A MDR EVENT.

Description of Event or Problem · 0

EVENT DESCRIPTION PATIENT REPORTED THAT THE MONITOR IS GIVING INACCURATE READINGS, AND ALSO CAUSE BRUISING DUE TO THE CUFF BEING TOO SMALL. MANUFACTURER NARRATIVE (PROVIDED BY LIVONGO) IT WAS DETERMINED THAT THE INACCURATE READINGS AND BRUISING THAT THE PATIENT REPORTED WERE DUE TO THE CUFF BEING TOO SMALL FOR THE PATIENTS ARM.THE DEVICE ASSOCIATED WITH THIS IINCIDENT WAS NOT RETURNED FOR INVESTIGATION.SHOULD THE DEVICE BE RETURNED, A SUPPLEMENTAL REPORT WILL BE FILED TO DDOCUMENT THE RESULTS OF THE INVESTIGATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2228508 LIVONGO BLOOD PRESSURE MONITOR DXN GUANGDONG TRANSTEK MEDICAL ELECTRONICS CO. LTD HT900

Patients

Seq Age Sex Outcome Treatment
1 53 YR Male