FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II Y 24GAX0.75IN PRN/EC SLM

MDR report key: 20485444 · Received October 18, 2024

Report

Report Number
3002601200-2024-00529
Event Type
Malfunction
Date Received
October 18, 2024
Date of Event
September 4, 2024
Report Date
October 22, 2024
Manufacturer
BD SUZHOU (MDS)
Product Code
FOZ
UDI-DI
00382903830336
PMA / PMN Number
UNK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

1. DHR/BHR REVIEW LOT#4052073. 1-THE PRODUCTS OF THIS BATCH WERE ASSEMBLED AT INTIMA II AUTO LINE 4 IN MARCH 2024, AND PACKAGED AT R240 PACKAGE LINE IN MARCH 2024. WORK ORDER QUANTITY WAS (B)(4) EA. 2-REVIEW THE IN-PROCESS TEST REPORTS AND OUTGOING TEST REPORTS, AND ALL TEST RESULTS MEET THE PRODUCT SPECIFICATIONS. 3-REVIEW THE PRODUCTION RECORDS WITH NO NONCONFORMANCE, DEVIATION OR REWORK ACTIVITIES. 2. NO DEFECTIVE SAMPLES AND PHOTOS HAVE BEEN RECEIVED FOR THE COMPLAINT. 3. CHECK THE RETAINED SAMPLES OF THIS BATCH, NO FOREIGN MATTERS ARE FOUND. PLEASE SEE ATTACHMENT FOR THE INSPECTION REPORT. 4. NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS REGARDING THIS BATCH OF PRODUCTS. CONCLUSION(S): NO ABNORMALITIES ARE FOUND IN THE PROCESS AND RETAINED SAMPLES, AND NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS ABOUT THIS BATCH OF PRODUCTS. AS THE STATUS AND COMPOSITION OF THE FOREIGN MATTER IN THE INDWELLING NEEDLE CANNOT BE CONFIRMED, THE ROOT CAUSE OF THIS DEFECT CANNOT BE DETERMINED. THE PLANT WILL CONTINUE TO PAY ATTENTION TO AND MONITOR SUCH DEFECT.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INTIMA-II Y 24GAX0.75IN PRN/EC SLM FOREIGN MATTER. ON (B)(6) 2024, A PATIENT WAS GIVEN AN INFUSION USING AN INDWELLING NEEDLE AND A FOREIGN OBJECT WAS FOUND IN THE INDWELLING NEEDLE, WHICH WAS LATER REPLACED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2229271 BD INTIMA-II Y 24GAX0.75IN PRN/EC SLM INTRAVASCULAR CATHETER FOZ BD SUZHOU (MDS) 4052073 00382903830336

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown