OSSEOTITE TAPERED IMPLANT 4 X 8.5MM
Report
- Report Number
- 0001038806-2024-02355
- Event Type
- Injury
- Date Received
- October 18, 2024
- Date of Event
- September 16, 2024
- Report Date
- October 17, 2024
- Manufacturer
- BIOMET 3I
- Product Code
- DZE
- PMA / PMN Number
- K063286
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
ZIMVIE COMPLAINT NUMBER (B)(4). D4: UNIQUE IDENTIFIER (UDI) NUMBER NOT AVAILABLE. D10. CONCOMITANT MEDICAL PRODUCTS NT485, OSSEOTITE TAPERED IMPLANT 4 X 8.5MM LOT 181351. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.
DURING A REVIEW AND RADIOLOGICAL CONTROL, SIGNIFICANT BONE LOSS AND INFLAMMATORY TISSUE AROUND THE IMPLANTS WERE DETECTED. DENTIST REPORTED A PERIIMPLANTITIS AT TOOTH SITE #36 AND #37. NO IMPACT ON THE PATIENT REPORTED. IMPLANTS WERE REMOVED. BONE DENSITY: II.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2208923 | OSSEOTITE TAPERED IMPLANT 4 X 8.5MM | DENTAL IMPLANT | DZE | BIOMET 3I | 181351 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Male | DENTAL IMPLANT SEE H10 NARRATIVE |