PUMP 1886 780G OUS BLE PUMP MG/DL
Report
- Report Number
- 2032227-2024-256606
- Event Type
- Malfunction
- Date Received
- October 18, 2024
- Date of Event
- October 1, 2024
- Report Date
- December 13, 2024
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000521554
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
PUMP WAS RECEIVED WITH A CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. UNABLE TO PERFORM THE FUNCTIONAL TESTS, INCLUDING THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST AND DISPLACEMENT TEST DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. SUCCESSFULLY DOWNLOADED PUMP TRACES AND HISTORY FILE USING THUMP. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM AND PUMP ERROR 35 ALARM CONFIRMED IN THE FORMATTED HISTORY FILE ON (B)(6) 2024 11:05:00.000 AND (B)(6) 2024 11:15:00.000. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND CORROSION ON THE PCBA 1, PCBA 2 AND FORCE SENSOR. NO CORROSION OR MOISTURE DAMAGE FOUND ON THE MOTOR AND VIBRATOR ASSEMBLY NOTED. PLEASE SEE BELOW FOR PUMP ERROR(S)/ALARM(S) NOTED 2 DAYS PRIOR TO THE EVENT DATE 01-OCT-2024 IN THE FORMATTED HISTORY FILE. LOSTSENSOR1ALERT (780) WAS FOUND ON: (B)(6) 2024 04:41:00.000 (B)(6) 2024 04:51:00.000 INSERT BATTERY ALARM WAS FOUND ON: (B)(6) 2024 11:28:51.000 (B)(6) 2024 11:35:17.000 FAILED BATTERY ALERT/BATTERY FAILED ALARM WAS FOUND ON: (B)(6) 2024 11:29:37.000 POWER MANAGEMENT GRAPH WAS SUCCESSFULLY GENERATED. THE POWER MANAGEMENT TOOL CONFIRMED THE UNLOADED VOLTAGE (UL VLITH) AND LOADED VOLTAGE (LOADED VLITH) WERE WITHIN SPEC RANGE. INSERT BATTERY ALARM WAS EXPECTED SINCE THE BATTERY WAS REMOVED FROM THE PUMP. UPON CHECKING ON THE POWER DATA/DETAIL TRACE FILE, FAILED BATTERY ALERT/BATTERY FAILED ALARM WAS EXPECTED DUE TO PUMP BATTERY DOES NOT HAVE ENOUGH POWER. THE CUSTOMER HAD USED A NO POWER/DEPLETED BATTERY. UNABLE TO TEST FOR LOST SENSOR ALERT AND FAILED BATTERY ALERT/BATTERY FAILED ALARM DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. LOST SENSOR ALERT AND FAILED BATTERY ALERT/BATTERY FAILED ALARM WERE UNKNOWN. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: A CRACKED KEYPAD OVERLAY AND A SCRATCHED CASE. UNABLE TO PERFORM THE REQUIRED TESTING DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM CONFIRMED DUE TO FATAL ALARM PUMP ERROR 35. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM AND PUMP ERROR 35 ALARM CONFIRMED DUE TO CORROSION ON THE PCBA 1, PCBA 2 AND FORCE SENSOR. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. THE REPORTED DEVICE IS NOT MARKETED IN THE UNITED STATES, BUT IT IS A SAME/SIMILAR DEVICE TO ONE THAT IS MARKETED INSIDE THE UNITED STATES. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED PUMP ERROR 35 (BROKEN FORCE SENSOR WAS DETECTED DURING SEATING OR REGULAR DELIVERY). THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1886. CUSTOMER REPORTED A CRITICAL PUMP ERROR OTHER THAN THE OPEN BOOK IMAGE ERROR OR CRITICAL PUMP ERROR 24. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE TO USE THE INSULIN PUMP. MMT-1886 WAS REQUESTED AND CUSTOMER RESPONSE WAS THE DEVICE WILL BE RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1716925 | PUMP 1886 780G OUS BLE PUMP MG/DL | AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1886 | NG3321590H | 000000763000521554 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 10 YR | Unknown |