FDA Adverse Event Malfunction Summary report: N

SPRINT FIDELIS

MDR report key: 2047717 · Received April 11, 2011

Report

Report Number
2649622-2011-04641
Event Type
Malfunction
Date Received
April 11, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S029
Removal / Correction Number
Z-0067-0070-2008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE SPRINT FIDELIS LEAD MODEL IS INCLUDED IN A FIELD ADVISORY. EVALUATION SUMMARY: (B)(4): THE ACTUAL LEAD WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. 9 - VENTRICULAR NST (NON-SUSTAINED TACHYCARDIA) <=210 MS AVERAGE V-CYCLE BETWEEN (B)(4)-2010 19:31:37 AND (B)(4)-2011 15:35:52. 1 - PATIENT ALERT FOR OUT OF TOLERANCE SUBTHRESHOLD IMPEDANCE ON (B)(4)-2010 19:36:01.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WERE SENSING DIFFICULTIES ON THE RIGHT VENTRICULAR LEAD. THERE WAS INTERMITTENT UNDERSENSING AND T WAVE OVERSENSING. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT FIDELIS IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6949 ASKU

Patients

Seq Age Sex Outcome Treatment
1 66 YR Other (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4076 IMPLANTABLE PACING LEAD