FDA Adverse Event Malfunction Summary report: N

PLUM 360 DRIVER NEW

MDR report key: 20474933 · Received October 17, 2024

Report

Report Number
9615050-2024-00766
Event Type
Malfunction
Date Received
October 17, 2024
Date of Event
September 18, 2024
Report Date
November 1, 2024
Manufacturer
ICU MEDICAL COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K141789
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

TESTS: SELF-TEST AND VISUAL INSPECTION. SELF-TEST FAILED DUE TO NOT TURNING ON AND SMOKE FROM THE DRIVER MOTOR PRINTED WIRING ASSEMBLY (PWA) BOARD. VISUAL INSPECTION PERFORMED AND FOUND TWO COMPONENTS WERE BURNT ON THE DRIVER PWA BOARD AND ONE COMPONENT WAS BURNT ON THE POWER SUPPLY PWA BOARD. THE CUSTOMER COMPLAINT WAS CONFIRMED THAT THE DEVICE DID NOT TURN ON AND SMOKE FROM THE DRIVER BOARD. THE PROBABLE CAUSE WAS A BURNT POWER SUPPLY PWA BOARD AND A BURNT DRIVER MOTOR PWA BOARD. FURTHERMORE, THE SERIAL NUMBER ON THE POWER SUPPLY PWA BOARD: (B)(6), DIDN'T MATCH THE ONE INSTALL BASE RECORDS, AND THE TAB ON THE MAIN CHASSIS WAS DAMAGED.

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED FOR EVALUATION; HOWEVER, TESTING HAS NOT YET BEEN COMPLETED.

Description of Event or Problem · 0

IT WAS REPORTED THAT A PLUM 360 PUMP BEGAN TO SMOKE FOLLOWING THE INSTALLATION OF A NEW POWER SUPPLY. THIS WAS DURING SET-UP. THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1276826 PLUM 360 DRIVER NEW PUMP, INFUSION FRN ICU MEDICAL COSTA RICA LTD.

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown