PLUM 360 DRIVER NEW
Report
- Report Number
- 9615050-2024-00766
- Event Type
- Malfunction
- Date Received
- October 17, 2024
- Date of Event
- September 18, 2024
- Report Date
- November 1, 2024
- Manufacturer
- ICU MEDICAL COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K141789
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
TESTS: SELF-TEST AND VISUAL INSPECTION. SELF-TEST FAILED DUE TO NOT TURNING ON AND SMOKE FROM THE DRIVER MOTOR PRINTED WIRING ASSEMBLY (PWA) BOARD. VISUAL INSPECTION PERFORMED AND FOUND TWO COMPONENTS WERE BURNT ON THE DRIVER PWA BOARD AND ONE COMPONENT WAS BURNT ON THE POWER SUPPLY PWA BOARD. THE CUSTOMER COMPLAINT WAS CONFIRMED THAT THE DEVICE DID NOT TURN ON AND SMOKE FROM THE DRIVER BOARD. THE PROBABLE CAUSE WAS A BURNT POWER SUPPLY PWA BOARD AND A BURNT DRIVER MOTOR PWA BOARD. FURTHERMORE, THE SERIAL NUMBER ON THE POWER SUPPLY PWA BOARD: (B)(6), DIDN'T MATCH THE ONE INSTALL BASE RECORDS, AND THE TAB ON THE MAIN CHASSIS WAS DAMAGED.
THE DEVICE WAS RETURNED FOR EVALUATION; HOWEVER, TESTING HAS NOT YET BEEN COMPLETED.
IT WAS REPORTED THAT A PLUM 360 PUMP BEGAN TO SMOKE FOLLOWING THE INSTALLATION OF A NEW POWER SUPPLY. THIS WAS DURING SET-UP. THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1276826 | PLUM 360 DRIVER NEW | PUMP, INFUSION | FRN | ICU MEDICAL COSTA RICA LTD. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |