VS3 IRIDIUM - LIGHT INTEGRATOR (BEAMCOMBINER)
Report
- Report Number
- 9616637-2024-00008
- Event Type
- Malfunction
- Date Received
- October 17, 2024
- Date of Event
- June 28, 2021
- Report Date
- October 17, 2024
- Manufacturer
- VISIONSENSE LTD.
- Product Code
- OWN
- UDI-DI
- 10884521817555
- PMA / PMN Number
- K150018
- Removal / Correction Number
- Z-0283-2023
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DA
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BEING SUBMITTED AS PART OF REMEDIATION ACTIVITIES ASSOCIATED WITH CAPA # (B)(4), WHICH IS ADDRESSING MDR REPORTING GU IDANCE FROM FDAS 1995 PREAMBLE, WHICH ENSURES COMPLAINTS RELATED TO CORRECTIONS AND RECALLS WHICH WERE PREVIOUSLY REPORTED TO FDA UNDER 21 CFR 806 ARE NOW REPORTED AS MDRS; THIS IS NOT A NEW MALFUNCTION/EVENT. H3 EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. VISUAL INSPECTION NOTED THE UNIT HAD A CABLE/WIRE FAILURE. IT WAS REPORTED THAT THE LIGHT WAS NOT CLEAR OR WAS DIM. THE REPORTED ISSUE WAS CONFIRMED. THE PRODUCT ANALYSIS NOTED EVIDENCE THAT THE DEVICE WAS NOT USED AS INTENDED. THIS ISSUE MAY OCCUR DUE TO MISHANDLING OR IMPROPER TRANSPORTATION. INTERNAL PROCESS IMPROVEMENTS HAVE BEEN INITIATED TO MITIGATE THIS ISSUE. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
ACCORDING TO THE REPORTER, DURING USE, THE UNIT LASER HAD NO OUTPUT FROM BEAM COMBINER- FIBER PINCHED. THERE WAS NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1287324 | VS3 IRIDIUM - LIGHT INTEGRATOR (BEAMCOMBINER) | CONFOCAL OPTICAL IMAGING | OWN | VISIONSENSE LTD. | 161-0001 | 10884521817555 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |