FDA Adverse Event Injury Summary report: N

MENTOR MEMORYGEL BREAST IMPLANT

MDR report key: 20473025 · Received October 17, 2024

Report

Report Number
1645337-2024-12212
Event Type
Injury
Date Received
October 17, 2024
Date of Event
July 16, 2024
Manufacturer
MENTOR TEXAS
Product Code
FTR
UDI-DI
00081317000464
PMA / PMN Number
P030053
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

SUSPECT DEVICE RECEIVED ON NOVEMBER 11, 2024. DEVICE EVALUATION COMPLETED ON NOVEMBER 14, 2024: THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED A VISUAL INSPECTION, MICROSCOPIC EXAMINATION, AND THICKNESS MEASUREMENT OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE SMOOTH HPG, 300CC BREAST IMPLANT WAS FOUND TO BE RUPTURED AND RECEIVED IN (2) TWO PARTS. MICROSCOPIC EXAMINATION WAS PERFORMED, AND THE CAUSE OF THE TEAR COULD NOT BE IDENTIFIED. THEREFORE A THICKNESS MEASUREMENT WAS CONDUCTED ON THE SHELL AT THE TEAR, AND THE THICKNESS WAS WITHIN MANUFACTURING SPECIFICATIONS. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. ALTHOUGH NO CONCLUSION COULD BE REACHED ON THE CAUSE OF THE REPORTED EVENT, THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING CAUTION: RUPTURE CAN OCCUR AT ANY TIME AFTER IMPLANTATION, BUT IT IS MORE LIKELY TO OCCUR THE LONGER THE IMPLANT IS IMPLANTED. THE FOLLOWING THINGS MAY CAUSE IMPLANT TO RUPTURE: DAMAGE BY SURGICAL INSTRUMENTS; STRESSING THE IMPLANT DURING IMPLANTATION AND WEAKENING IT; FOLDING OR WRINKLING OF THE IMPLANT SHELL; EXCESSIVE FORCE TO THE CHEST (E.G. DURING CLOSED CAPSULOTOMY); TRAUMA; COMPRESSION DURING MAMMOGRAPHIC IMAGING; AND SEVERE CAPSULAR CONTRACTURE. BREAST IMPLANTS MAY ALSO SIMPLY WEAR OUT OVER TIME. A SECOND PRODUCT WAS RECEIVED (LOT-6803159). NO ADVERSE EVENTS WERE REPORTED FOR THIS CONCOMITANT (CONTRALATERAL) DEVICE, THEREFORE NO FURTHER ANALYSIS IS REQUIRED. AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: SILENT RUPTURE . MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION RECEIVED ON (B)(6) 2024 INDICATED THAT THE PATIENT UNDERWENT BILATERAL REPLACEMENTS AS FOLLOW: R) REF (B)(4) S/N (B)(6), L) REF (B)(4) S/N (B)(6). MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A FEMALE WHO UNDERWENT A BREAST AUGMENTATION PRIMARY WITH MENTOR MEMORYGEL, 300CC SILICONE PROSTHESIS EXPERIENCED RIGHT-SIDED SILENT RUPTURE POSTOPERATIVE. THE ISSUE WAS DIAGNOSED THROUGH PHYSICAL AND CONFIRMED BY MRI EXAMINATION. AS A RESULT, PATIENT UNDERWENT REMOVAL ON (B)(6) 2024.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1202625 MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS 6803159 00081317000464

Patients

Seq Age Sex Outcome Treatment
1 47 YR Female Required Intervention