FDA Adverse Event Malfunction Summary report: N

DAVINCI X

MDR report key: 20467871 · Received October 16, 2024

Report

Report Number
2955842-2024-20676
Event Type
Malfunction
Date Received
October 16, 2024
Date of Event
September 24, 2024
Report Date
September 24, 2024
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874115404
PMA / PMN Number
K171294
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. THE FSE WAS ABLE TO REPRODUCE THE REPORTED COMPLAINT AND REPLACED THE UNIVERSAL SURGICAL MANIPULATOR (USM) TO RESOLVE THE ISSUE. THE SYSTEM WAS VERIFIED AND READY TO USE. ISI HAS ISSUED A RETURN MATERIAL AUTHORIZATION (RMA) TO HAVE THE USM RETURNED. HOWEVER, ISI HAS NOT RECEIVED THE PRODUCT INVOLVED WITH THE ALLEGED ISSUE TO PERFORM FAILURE ANALYSIS.

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. THE FSE REPLACED THE UNIVERSAL SURGICAL MANIPULATOR (USM) TO RESOLVE THE ISSUE. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. ISI DID RECEIVE A DA VINCI PRODUCT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS. THE USM WAS ANALYZED AND THE REPORTED FAILURE WAS NOT CONFIRMED AS HAVING OCCURRED IN THE FIELD AND NOT REPLICATED. LOGS RECORDED ERROR 40100 COUPLED WITH ERRORS 32114 AND 32097 WHICH POINTS TO THE CLOCK SPRING. UNIT WAS TESTED ON AN IN-HOUSE SYSTEM IN NORMAL MODE WITH NO TRIGGERED ERRORS. UNIT WAS ALSO TESTED ON A PFTP WHERE ALL TEST PASSED. AS A PRECAUTION, THE CLOCK SPRING ASSEMBLY WILL BE REPLACED. THE COMPLAINT WAS NOT CONFIRMED BY FAILURE ANALYSIS.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED PROSTATECTOMY SURGICAL PROCEDURE, THERE WAS REPEATED RECOVERABLE FAULT 40100 ON UNIVERSAL SURGICAL MANIPULATOR (USM) 1. USM 1 WAS RED AND WAS UNUSABLE. THE INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) REVIEWED THE LOGS AND FOUND ERROR 25730 POINTING TO MOTOR AXIS ERROR ON AXES CONTROLLER, ARM (ACA) ON USM 1. THE TSE GUIDED THE CALLER THROUGH HARD POWER CYCLE AND EMERGENCY POWER OFF (EPO) ON PATIENT SIDE CART (PSC). THE SYSTEM STARTED NORMALLY WITH NO ERRORS, BUT AS SOON AS USM 1 WAS MOVED, THE ERROR SHOWED UP AGAIN. THE TSE ASKED THE CALLER IF THE CUSTOMER COULD PROCEED WITH 3 USMS, WHICH WAS THE CASE. USM1 WAS DISABLED, AND SURGERY WAS RESUMED. THE PROCEDURE WAS COMPLETING AS PLANNED WITH NO REPORTED INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1033755 DAVINCI X PATIENT SIDE CART, 4-ARM NAY INTUITIVE SURGICAL, INC 380620-48 N/A 00886874115404

Patients

Seq Age Sex Outcome Treatment
1 NA Male DA VINCI INSTRUMENTS AND ACCESSORIES