FDA Adverse Event Injury Summary report: N

UNICEL® DXC 800 PRO SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 2046471 · Received April 8, 2011

Report

Report Number
2050012-2011-00990
Event Type
Injury
Date Received
April 8, 2011
Date of Event
March 8, 2011
Report Date
March 9, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SAMPLE TYPE FOR THIS ASSAY IS RANDOM URINE. NO SAMPLE ISSUES WERE NOTED. PER CUSTOMER, QC HAS BEEN WITHIN RANGE. NO SYSTEM ERRORS WERE NOTED. BCI FIELD SERVICE ENGINEER (FSE) PERFORMED THE FOLLOWING: INSPECTED INSTRUMENT AND FOUND CARTRIDGE CHEMISTRY (CC) SAMPLE MIXER ASSEMBLY HAD A WORN TIP. REPLACED CC SAMPLE AND REAGENT MIXER ASSEMBLY AND VERIFIED ALIGNMENTS. REPLACED DEGASSER FILTER AND CC SAMPLE WASH COLLAR. VERIFIED WASH DELIVERY INSIDE MIXER WASH CUPS FOR SAMPLE AND REAGENT. VERIFIED THAT DELIVERY OF WATER DURING ONE PRIME CYCLE FOR CC SAMPLE PROBE IS EXPECTED VALUE OF 35ML. VERIFIED THAT WASH SOLUTION IS MADE CORRECTLY BY INSTRUMENT BY VERIFYING K LEVEL OF SOLUTION WAS WITHIN RANGE. RAN MA CARRYOVER BY ALTERNATING SERUM SAMPLES WITH URINE MA SAMPLES. STILL GETTING CARRYOVER ON FIRST REP OF MA FOLLOWING A SERUM TEST IS ELEVATED BY 15 UNITS. REPLACED SAMPLE PROBE ASSEMBLY AND RERAN SERUM/URINE MA CARRYOVER. ALL MA RESULTS WERE WITHIN EXPECTED PRECISION AND NO CARRYOVER DETECTED. MA ELEVATED RESULTS ARE DUE TO SAMPLE PROBE CARRYOVER. THREE SAMPLE PROBES BEING SENT TO CUSTOMER FOR FUTURE TROUBLE SHOOTING AND REPLACEMENT IF NEEDED. FSE RECOMMENDED FLUSHING CC SAMPLE PROBE USING BCI PROCEDURE IF CARRYOVER IS SUSPECTED AS ANOTHER ALTERNATIVE TO REPLACEMENT. THE SAMPLE PROBE REPLACEMENT APPEARS TO HAVE RESOLVED THE ISSUE.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO ERRONEOUS HIGH MICRO ALBUMIN (MA) RESULTS GENERATED BY THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. THE ERRONEOUS RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. THE SAMPLES WERE REPEATED AND LOWER RESULTS WERE OBTAINED. NO AFFECT TO PATIENTS, THE INCORRECT RESULTS HAVE NOT BEEN REPORTED OUT OF THE LABORATORY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 800 PRO SYNCHRON ® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 800 PRO N/A

Patients

Seq Age Sex Outcome Treatment
1