UNICEL® DXC 800 PRO SYNCHRON ® CHEMISTRY ANALYZER
Report
- Report Number
- 2050012-2011-00989
- Event Type
- Injury
- Date Received
- April 8, 2011
- Date of Event
- March 7, 2011
- Report Date
- March 9, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
SAMPLE TYPE FOR THIS ASSAY IS RANDOM URINE. NO SAMPLE ISSUES WERE NOTED. PER CUSTOMER, QC HAS BEEN WITHIN RANGE. NO SYSTEM ERRORS WERE NOTED. BCI FIELD SERVICE ENGINEER (FSE) PERFORMED THE FOLLOWING: INSPECTED INSTRUMENT AND FOUND CARTRIDGE CHEMISTRY (CC) SAMPLE MIXER ASSEMBLY HAD A WORN TIP. REPLACED CC SAMPLE AND REAGENT MIXER ASSEMBLY AND VERIFIED ALIGNMENTS. REPLACED DEGASSER FILTER AND CC SAMPLE WASH COLLAR. VERIFIED WASH DELIVERY INSIDE MIXER WASH CUPS FOR SAMPLE AND REAGENT. VERIFIED THAT DELIVERY OF WATER DURING ONE PRIME CYCLE FOR CC SAMPLE PROBE IS EXPECTED VALUE OF 35ML. VERIFIED THAT WASH SOLUTION IS MADE CORRECTLY BY INSTRUMENT BY VERIFYING K LEVEL OF SOLUTION WAS WITHIN RANGE. RAN MA CARRYOVER BY ALTERNATING SERUM SAMPLES WITH URINE MA SAMPLES. STILL GETTING CARRYOVER ON FIRST REP OF MA FOLLOWING A SERUM TEST IS ELEVATED BY 15 UNITS. REPLACED SAMPLE PROBE ASSEMBLY AND RERAN SERUM/URINE MA CARRYOVER. ALL MA RESULTS WERE WITHIN EXPECTED PRECISION AND NO CARRYOVER DETECTED. MA ELEVATED RESULTS ARE DUE TO SAMPLE PROBE CARRYOVER. THREE SAMPLE PROBES BEING SENT TO CUSTOMER FOR FUTURE TROUBLE SHOOTING AND REPLACEMENT IF NEEDED. FSE RECOMMENDED FLUSHING CC SAMPLE PROBE USING BCI PROCEDURE IF CARRYOVER IS SUSPECTED AS ANOTHER ALTERNATIVE TO REPLACEMENT. THE SAMPLE PROBE REPLACEMENT APPEARS TO HAVE RESOLVED THE ISSUE.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO ERRONEOUS HIGH MICRO ALBUMIN (MA) RESULTS GENERATED BY THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. THE ERRONEOUS RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. THE SAMPLES WERE REPEATED AND LOWER RESULTS WERE OBTAINED. NO AFFECT TO PATIENTS, THE INCORRECT RESULTS HAVE NOT BEEN REPORTED OUT OF THE LABORATORY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON ® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 800 PRO | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |