LEFTY TIGHTY LAG SCREW, Ø10.5MMX100MM
Report
- Report Number
- 1220246-2024-08166
- Event Type
- Injury
- Date Received
- October 16, 2024
- Date of Event
- August 27, 2024
- Report Date
- June 10, 2025
- Manufacturer
- ARTHREX, INC.
- Product Code
- HSB
- UDI-DI
- 00848665036558
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- 501
Narratives
INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED UPON AVAILABILITY OF ADDITIONAL INFORMATION.
ADDITIONAL INFORMATION: G3, H3, H6 BASED ON THE INFORMATION PROVIDED WHICH MAY INCLUDE THE DEVICE (IF AVAILABLE AND RETURNED), PICTURES, VIDEOS, EVENT DESCRIPTION, AND ANY ADDITIONAL INFORMATION FROM THE FIELD, ARTHREX WAS ABLE TO CONCLUDE A MOST LIKELY CAUSE. THE MOST LIKELY CAUSE FOR THE REPORTED FAILURE CAN BE ATTRIBUTED TO USER ERROR OF THE DEVICE DUE TO MISALIGNED INSERTION.
ON 9/10/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT A 0857-100 CAPTURING ROD AND 1192-100 LEFTY TIGHTY LAG SCREW GOT STUCK TOGETHER AND COULD NOT BE TAKEN APART. THIS WAS DISCOVERED DURING A TROCH NAIL PROCEDURE ON (B)(6) 2024. THE CASE WAS COMPLETED USING A NEW 0857-100 CAPTURING ROD AND 1192-100 LEFTY TIGHTLY LAG SCREW. ADDITIONAL INFORMATION RECEIVED ON 9/27/2024: IT WAS NOTICED AFTER THE SURGERY THAT THE 0857-100 CAPTURING ROD AND 1192-100 LEFTY TIGHTY LAG SCREW GOT STUCK TOGETHER. ADDITIONAL INFORMATION RECEIVED ON 10/4/2024: THIS OCCURRED DURING A REVISION SURGERY. THE SCREW HAD COLLAPSED ALL THE WAY AND ENDED UP CUTTING THROUGH THE HEAD. ALL PRODUCTS WERE REMOVED, AND THE CASE WAS COMPLETED AS A TOTAL HIP PROCEDURE. THE ORIGINAL SURGERY WAS PERFORMED ON (B)(6) 2024. BOTH SURGERIES WERE PERFORMED AT THE SAME FACILITY AND BY THE SAME SURGEON.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2557122 | LEFTY TIGHTY LAG SCREW, Ø10.5MMX100MM | INTRAMED FIXATN ROD AND ACCESS | HSB | ARTHREX, INC. | LEFTY TIGHTY LAG SCREW, Ø10.5MMX100MM | 230094 | 00848665036558 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Other |