COBAS E 411 ANALYZER (RACK SYSTEM)
Report
- Report Number
- 1823260-2024-02990
- Event Type
- Malfunction
- Date Received
- October 16, 2024
- Date of Event
- September 13, 2024
- Report Date
- December 30, 2024
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- UDI-DI
- 04015630924721
- PMA / PMN Number
- K961481
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SF
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
CORRECTION: THE PCT REAGENT LOT NUMBER WAS 755916. THE FT4 REAGENT LOT NUMBER WAS 767235. THE INVESTIGATION IS ONGOING.
THE FIELD SERVICE ENGINEER (FSE) ADJUSTED THE RACK LOADER SAMPLE POSITION. THE INSTRUMENT WAS VERIFIED AND ALL CHECKS WERE PERFORMED SUCCESSFULLY. THE ROOT CAUSE WAS DUE TO INSUFFICIENT COMPONENT ADJUSTMENT. THE SERVICE ACTIONS PERFORMED BY THE FSE RESOLVED THE ISSUE. NO FURTHER ISSUES WERE REPORTED AFTERWARD.
THE PCT REAGENT LOT NUMBER WAS 120820. THE EXPIRATION DATE WAS NOT PROVIDED. THE FT4 REAGENT LOT NUMBER WAS 105782. THE EXPIRATION DATE WAS NOT PROVIDED. ON 13-SEP-2024 AND 14-SEP-2024, THE FSE FOUND THAT THE RACK LOADER WAS SLIGHTLY SEPARATED FROM THE INSTRUMENT COMPROMISING THE RACK SAMPLING POSITION ADJUSTMENT. HE POSITIONED THE RACK LOADER BACK INTO PLACE. HE THEN RAN TEST SAMPLES AND THEY WERE ACCEPTABLE. HE CONFIRMED THAT THE INSTRUMENT WAS PERFORMING BACK WITHIN SPECIFICATIONS. NO FURTHER ISSUES WERE OBSERVED. THE CUSTOMER MENTIONED THAT THEY RECEIVED LLD ERRORS AGAIN. ON (B)(6) 2024, THE FSE REPLACED THE SAMPLE/REAGENT PULLEY TUBING AND VERIFIED THAT THE LLD PRINTED CIRCUIT BOARD (PCB) WAS POSITIONED PROPERLY. HE ALSO FOUND THAT THE LLD VOLTAGE ON THE SAMPLE/REAGENT PROBE WAS SLIGHTLY LOWER THAN EXPECTED AND HE ADJUSTED IT TO 0.74. THE FSE THEN CHECKED ALL ADJUSTMENTS AND THEY WERE ALL WITHIN SPECIFICATIONS. HE PERFORMED CALIBRATION, QC, AND RAN TEST SAMPLES AND THEY WERE ALL ACCEPTABLE. THE INVESTIGATION IS ONGOING.
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT SERUM/PLASMA SAMPLE TESTED FOR PROCALCITONIN (PCT) AND 1 PATIENT SERUM/PLASMA SAMPLE TESTED FOR FT4 ON A COBAS E411 IMMUNOASSAY ANALYZER (RACK SYSTEM). SAMPLE 1 (PATIENT 1) WAS TESTED FOR PCT: INITIAL RESULT: 0.104 NG/ML. REPEAT RESULT: 17.03 NG/ML. SAMPLE 2 (PATIENT 2) WAS TESTED FOR FT4: INITIAL RESULT: 51.51 PMOL/L. THIS RESULT WAS CONSIDERED ABOVE THE EXPECTED RANGE (11.90 PMOL/L TO 21.60 PMOL/L). THE SAMPLE WAS THEN REPEATED ON 14-SEP-2024 AND THE REPEAT RESULT WAS 17.82 PMOL/L. THE CUSTOMER MENTIONED THAT THEY RECEIVED LIQUID LEVEL DETECTION (LLD) ERRORS ON (B)(6) 2024. THE REPEAT RESULTS WERE OBTAINED AFTER THE FIELD SERVICE ENGINEER (FSE) VISITED THE CUSTOMER'S SITE ON (B)(6) 2024 AND PERFORMED TROUBLESHOOTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2558935 | COBAS E 411 ANALYZER (RACK SYSTEM) | IMMUNOCHEMISTRY ANALYZER | JJE | ROCHE DIAGNOSTICS | 04015630924721 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |