FDA Adverse Event Malfunction Summary report: N

TINA-QUANT ¿2-MICROGLOBULIN

MDR report key: 20461807 · Received October 16, 2024

Report

Report Number
1823260-2024-02986
Event Type
Malfunction
Date Received
October 16, 2024
Date of Event
September 18, 2024
Report Date
November 26, 2024
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JZG
UDI-DI
07613336119990
PMA / PMN Number
K980724
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE CALIBRATION AND QC WERE WITHIN SPECIFICATIONS. THE FIELD SERVICE REPRESENTATIVE CORRECTLY PROGRAMMED ALL SPECIAL WASH FUNCTIONS BEFORE PIPETTING. CALIBRATIONS AND QCS WERE PERFORMED AND THE RESULTS WERE WITHIN SPECIFICATIONS. NO FURTHER ISSUES WERE REPORTED BY THE CUSTOMER. THE INVESTIGATION DETERMINED THE SERVICE ACTIONS RESOLVED THE ISSUE. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM.

Additional Manufacturer Narrative · 0

THE ANALYZER'S SERIAL NUMBER IS (B)(6). THE INVESTIGATION IS ONGOING.

Description of Event or Problem · 0

THERE WAS AN ALLEGATION OF QUESTIONABLE TINA-QUANT 2-MICROGLOBULIN RESULTS FOR 2 PATIENT SAMPLES ON A COBAS 8000 C 502 MODULE. PATIENT 1: THE INITIAL RESULT WAS INVALIDATED DUE TO HAVING A DATA FLAG. THE SAMPLE WAS AUTOMATICALLY DILUTED AND THE RESULT WAS 6.4 MG/L. THE SAMPLE WAS MANUALLY DILUTED AND THE RESULT WAS 16.8 MG/L. THIS RESULT WAS BELIEVED TO BE CORRECT. PATIENT 2: THE INITIAL RESULT WAS INVALIDATED DUE TO HAVING A DATA FLAG. THE SAMPLE WAS AUTOMATICALLY DILUTED AND THE RESULT WAS 7.2 MG/L. THE SAMPLE WAS MANUALLY DILUTED AND THE RESULT WAS 16.8 MG/L. THIS RESULT WAS BELIEVED TO BE CORRECT. THE SAMPLES WERE REPEATED DUE TO THE INITIAL RESULT HAVING A DATA FLAG, AND HIGHER RESULTS WERE EXPECTED. THE QUESTIONABLE RESULTS WERE NOT REPORTED OUTSIDE THE LABORATORY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2582125 TINA-QUANT ¿2-MICROGLOBULIN BETA-2-MICROGLOBULIN IMMUNOLOGICAL TEST SYSTEM JZG ROCHE DIAGNOSTICS 76651501 07613336119990

Patients

Seq Age Sex Outcome Treatment
1 70 YR Male