COMPRESSION TOOL
Report
- Report Number
- 2032499-2011-00076
- Event Type
- Malfunction
- Date Received
- February 28, 2011
- Date of Event
- March 26, 2009
- Report Date
- March 27, 2009
- Manufacturer
- INTERVENTIONAL SPINE, INC.
- Product Code
- HWN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
ROOT CAUSE: THE PRONGS WERE SLIGHTLY DEFORMED WITH USE. WHEN USED IN THIS COMPLAINT, THE DEFORMED PRONGS CAUSED THE PIN TO BE IMPINGED BEFORE GRABBING THE PIN DURING ACTUATION. THIS WAS MOST PROBABLY CAUSED BY ROUTINE WEAR AND TEAR. THE INSTRUCTIONS FOR USE CLEARLY STATE THAT ALL INSTRUMENTS SHOULD BE INSPECTED FOR DAMAGE DUE TO WEAR AND TEAR PRIOR TO USE. INITIALLY IT WAS DETERMINED THE EVENT WAS NOT REPORTABLE. ON A SUBSEQUENT RE-REVIEW, IT WAS DETERMINED THIS EVENT SHOULD HAVE BEEN REPORTED. DEVICE EVALUATION: RETURNED DEVICE WAS INSPECTED VISUALLY. OUTER SHEATH WAS REMOVED, REVEALING INNER MECHANISMS. THE PRONGS, WHICH GRAB THE IMPLANT, WERE SLIGHTLY DEFORMED. HOWEVER, THE INSTRUMENT PROVED TO BE FUNCTIONAL WHEN BENCH-TESTED. A SAMPLE IMPLANT WAS COMPRESSED EASILY USING THE RETURNED DEVICE. DURING A CASE, THE ANGLE AT WHICH THE INSTRUMENT IS HELD MAY CAUSE THE DEVICE TO MALFUNCTION.
COMPRESSION TOOL WAS NOT GRABBING PULL PIN AND ALLOWING COMPRESSION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COMPRESSION TOOL | HWN | INTERVENTIONAL SPINE, INC. | 9049-00 | 112107-41 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |