FDA Adverse Event Malfunction Summary report: N

IMMAGE® MICROALBUMIN (MA) REAGENT

MDR report key: 2044583 · Received April 7, 2011

Report

Report Number
2050012-2011-01041
Event Type
Malfunction
Date Received
April 7, 2011
Date of Event
March 9, 2011
Report Date
March 9, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JIQ
PMA / PMN Number
K965035
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO SAMPLE INFORMATION WAS PROVIDED. THE SAMPLE TYPE FOR MICROALBUMIN IS USUALLY RANDOM URINE. CUSTOMER NOTED THAT QC HAS BEEN UNACCEPTABLE. CUSTOMER DID NOT NOTE ANY SYSTEM ERRORS OR ISSUES. IT IS UNKNOWN WHETHER A NEW LOT OR CARTRIDGE OF REAGENT RESOLVED THE ISSUE. SERVICE WAS NOT NEEDED FOR THIS EVENT, AS THIS IS A REAGENT ISSUE.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT THE RESULTS OF CONTROLS AND PATIENT SPECIMENS ARE UNSTABLE FOR THE MICROALBUMIN REAGENT. THE CUSTOMER ALSO STATED THAT THE MICROALBUMIN REAGENT GENERATED DIFFERENT RESULTS WHEN LOADED ON DIFFERENT CARTRIDGES. NO SAMPLE INFORMATION WAS PROVIDED EXCEPT THAT SAMPLE RESULTS WERE DIFFERENT WITH DIFFERENT LOTS OR CARTRIDGES OF REAGENT. NO AFFECT TO PATIENTS WERE REPORTED, AS THE RESULTS WERE NOT REPORTED OUT OF THE LABORATORY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 IMMAGE® MICROALBUMIN (MA) REAGENT MICROALBUMIN (MA) REAGENT JIQ BECKMAN COULTER INC. N/A M010304

Patients

Seq Age Sex Outcome Treatment
1