FDA Adverse Event Malfunction Summary report: N

BD NEXIVA SP WITH MAXZERO

MDR report key: 20440467 · Received October 14, 2024

Report

Report Number
1710034-2024-01142
Event Type
Malfunction
Date Received
October 14, 2024
Date of Event
September 23, 2024
Report Date
December 2, 2024
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903835577
PMA / PMN Number
K183399
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED. B3. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD.

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: THE COMPLAINT THAT THE NEEDLE WAS DIFFICULT TO REMOVE WAS CONFIRMED AND THE CAUSE APPEARED TO BE SUPPLIER RELATED. REPRESENTATIVE 20G NEXIVA UNITS FROM LOT# 4152107 WERE PROVIDED FOR INVESTIGATION. TWO SAMPLES WERE FOUND WITH RESISTANCE BETWEEN THE NEEDLE AND SAFETY MECHANISM. THE OUTSIDE DIAMETER OF THE NEEDLE WAS WITHIN SPEC. THE INSIDE DIAMETER OF THE SAFETY MECHANISM COMPONENT WAS BELOW SPECIFICATION, WHICH CREATED FRICTION AS THE NEEDLE WITHDRAWN THROUGH THE SAFETY MECHANISM. THE APPROPRIATE PERSONNEL WERE NOTIFIED OF THIS COMPLAINT. A REVIEW OF OUR RISK MANAGEMENT DOCUMENTATION WAS PERFORMED AND INDICATES THAT THE POTENTIAL RISK OF THE REPORTED EVENT WAS ASSESSED APPROPRIATELY.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD NEXIVA 20 GA X 1.25IN SP WITH MAXZERO DIFFICULT TO DISENGAGE NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN ATTEMPTING TO REMOVE THE NEEDLE, THEY TEND TO GET STUCK AND ARE DIFFICULT TO EXTRACT. INJURIES OR ADVERSE EVENT: NO. (B)(6) 2024. THIS HAS HAPPENED THROUGHOUT THE MONTH OF SEPTEMBER; I DO NOT HAVE EXACT DATES.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1716431 BD NEXIVA SP WITH MAXZERO PERIPHERAL IV CATHETERS FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 4152107 00382903835577

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown