FDA Adverse Event Injury Summary report: N

PERCLOSE¿ PROSTYLE¿

MDR report key: 20439968 · Received October 14, 2024

Report

Report Number
2024168-2024-12072
Event Type
Injury
Date Received
October 14, 2024
Date of Event
September 17, 2024
Report Date
February 4, 2025
Manufacturer
ABBOTT VASCULAR INC.
Product Code
MGB
UDI-DI
08717648235184
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT. THE ADDITIONAL DEVICES REFERENCED IN B5 ARE FILED UNDER SEPARATE MEDWATCH REPORT NUMBERS.

Additional Manufacturer Narrative · 0

ANALYSIS WAS PERFORMED ON THE RETURNED DEVICE. THE REPORTED NEEDLE TO CUFF MISS WAS OBSERVED AS A LINK SEPARATION. PRODUCTION RECORD AND CORRECTIVE AND PREVENTIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. THE REPORTED DIFFICULTY AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR SUTURE/ LINK PULLED UNTIL IT BREAKS DUE TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING OF THE DEVICE. B5: DESCRIBE EVENT OR PROBLEM: UPDATED. D4: LOT # CORRECTED FROM 4071142 TO 4071941. D4: PRIMARY UDI NUMBER CORRECTED TO (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS WAS A VENOUS CLOSURE OF THE RIGHT COMMON FEMORAL VEIN USING A PROSTYLE DEVICE AFTER AN ABLATION INTERVENTIONAL PROCEDURE. REPORTEDLY, A CUFF MISS [SUTURE RETRIEVAL ISSUE] OCCURRED WITH 6 PROSTYLE DEVICES. A NEW PROSTYLE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Description of Event or Problem · 0

SUBSEQUENT TO THE PREVIOUSLY FILED REPORT, AN ADDITIONAL PLUNGER WAS RETURNED WHICH SHOWED SIGNS OF A CUFF MISS. THE ACCOUNT WAS UNABLE TO PROVIDE FURTHER INFORMATION REGARDING THE ADDITIONAL DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1715421 PERCLOSE¿ PROSTYLE¿ DEVICE, HEMOSTASIS, VASCULAR MGB ABBOTT VASCULAR INC. 12773-03 4071941 08717648235184

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention