FDA Adverse Event Malfunction Summary report: N

LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES

MDR report key: 2043614 · Received March 29, 2011

Report

Report Number
3015876-2011-00280
Event Type
Malfunction
Date Received
March 29, 2011
Date of Event
March 4, 2011
Report Date
March 4, 2011
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
MKJ
PMA / PMN Number
K991910
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND REPLACED THE POWER PCB ASSEMBLY. PROPER DEVICE OPERATION WAS OBSERVED THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE LOANER POOL FOR USE. PHYSIO FURTHER EVALUATED THE REMOVED POWER PCB ASSEMBLY AND DETERMINED THE CAUSE OF THE REPORTED FAILURE WAS DUE TO A SHORTED CAPACITOR, DESIGNATOR (B)(4).

Description of Event or Problem · 1

IT WAS OBSERVED THAT THE PHYSIO LOANER DEVICE WOULD NOT POWER ON WITH AC OR DC POWER. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES MKJ PHYSIO-CONTROL, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA