FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES
MDR report key: 2043614
·
Received March 29, 2011
Report
- Report Number
- 3015876-2011-00280
- Event Type
- Malfunction
- Date Received
- March 29, 2011
- Date of Event
- March 4, 2011
- Report Date
- March 4, 2011
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- MKJ
- PMA / PMN Number
- K991910
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND REPLACED THE POWER PCB ASSEMBLY. PROPER DEVICE OPERATION WAS OBSERVED THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE LOANER POOL FOR USE. PHYSIO FURTHER EVALUATED THE REMOVED POWER PCB ASSEMBLY AND DETERMINED THE CAUSE OF THE REPORTED FAILURE WAS DUE TO A SHORTED CAPACITOR, DESIGNATOR (B)(4).
Description of Event or Problem · 1
IT WAS OBSERVED THAT THE PHYSIO LOANER DEVICE WOULD NOT POWER ON WITH AC OR DC POWER. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES | MKJ | PHYSIO-CONTROL, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |