FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 2043570 · Received March 28, 2011

Report

Report Number
1644487-2011-00655
Event Type
Malfunction
Date Received
March 28, 2011
Date of Event
July 11, 2006
Report Date
March 25, 2011
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF PROGRAMMING HISTORY.

Description of Event or Problem · 1

DURING MFR REVIEW OF A VNS PT'S PROGRAMMING HISTORY, IT WAS NOTED A FAULTED SYSTEMS DIAGNOSTICS TEST OCCURRED ON (B)(6) 2006, WHICH CAUSED THE SETTINGS TO CHANGE TO SYSTEMS DIAGNOSTICS DEFAULT SETTINGS. THE CHANGE WAS NOTED UPON REINTERROGATION OF THE PT'S GENERATOR AND ALL SETTINGS WERE CORRECTED EXCEPT FOR THE OFF TIME, WHICH REMAINED AT 60 MINUTES. THE OFF TIME WAS NOT CORRECTED UNTIL (B)(6) 2006.