HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1423500-2011-04173
- Event Type
- Malfunction
- Date Received
- April 7, 2011
- Date of Event
- March 9, 2011
- Report Date
- March 17, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE SYSTEM ERROR 2240 WAS IDENTIFIED DURING A REVIEW OF A RETURNED HOMECHOICE DEVICE; THERE WAS NO REPORT FOR THIS ALARM CALLED IN BY THE CUSTOMER AND NO ALLEGATION OF A PRODUCT MALFUNCTION, THEREFORE, THE SAMPLE WAS NOT REQUESTED. NOT ENOUGH DATA IS AVAILABLE WITHIN THE COMPLAINT TO IDENTIFY ROOT CAUSE; THEREFORE, THE CAUSE WAS NOT DETERMINED. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
(B)(4). THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER, THEREFORE THE DEVICE WAS NOT REQUESTED FOR EVALUATION AND A BATCH REVIEW WILL NOT BE CONDUCTED.
A SYSTEM ERROR (SE) 2240 ALARM WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE DEVICE. THE SYSTEM ERROR OCCURRED ON (B)(6) 2011 DURING THE INITIAL DRAIN CYCLE. A SE 2240 IS A NON-RECOVERABLE ALARM THAT OCCURS WHEN THE DEVICE HAS DETECTED AIR BEING DRAWN CONTINUOUSLY INTO THE DISPOSABLE. IT IS UNKNOWN IF THIS ALARM OCCURRED DURING PATIENT THERAPY, HOWEVER NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. NO FURTHER INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR | HOMECHOICE |