FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 2043433
·
Received March 28, 2011
Report
- Report Number
- 1720753-2011-02919
- Event Type
- Malfunction
- Date Received
- March 28, 2011
- Date of Event
- February 16, 2011
- Report Date
- March 28, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE INSULATED GATE BIPOLAR TRANSISTOR AND A FUSE WERE REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE SYSTEM DISPLAYED A LOW KILOVOLTS ERROR CODE MESSAGE. NO REPORT OF PATIENT INJURY.