FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2043433 · Received March 28, 2011

Report

Report Number
1720753-2011-02919
Event Type
Malfunction
Date Received
March 28, 2011
Date of Event
February 16, 2011
Report Date
March 28, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE INSULATED GATE BIPOLAR TRANSISTOR AND A FUSE WERE REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM DISPLAYED A LOW KILOVOLTS ERROR CODE MESSAGE. NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1