MERIT CUSTOM FIT
Report
- Report Number
- 1721504-2011-00071
- Event Type
- Malfunction
- Date Received
- February 25, 2011
- Date of Event
- December 22, 2010
- Report Date
- February 3, 2011
- Manufacturer
- MERIT MEDICAL SYSTEMS, INC.
- Product Code
- DTL
- PMA / PMN Number
- K934123
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVAL: ONE USED SUSPECT DEVICE WAS RETURNED FOR EVAL. UPON VISUAL INSPECTION THE COMPLAINT WAS CONFIRMED. THE ROTATOR ON THE STOPCOCK WAS SEPARATED AT THE BOND BETWEEN THE COLLAR AND THE HOUSING. A REVIEW OF THE DEVICE HISTORY RECORD DID NOT REVEAL ANY EXCEPTION DOCUMENTS. THE COMPLAINT DATA BASE WAS REVIEWED AND FOUND NO SIMILAR COMPLAINTS FOR THIS LOT NUMBER. THE ROOT CAUSES OF THE FAILURE ARE ATTRIBUTED TO THE MFG BONDING PROCESS. CORRECTIVE ACTIONS HAVE BEEN IMPLEMENTED TO ADDRESS THIS TYPE OF FAILURE. THIS SUBASSEMBLY WAS BUILT PRIOR TO IMPLEMENTATION OF THE NOTED CORRECTIVE ACTIONS. EVAL METHOD: DEVICE HISTORY RECORD WAS REVIEWED, COMPLAINT DATA BASE WAS REVIEWED.
THE ROTATOR BROKE DURING USE. INJECTOR SETTINGS: FLOW 10 ML/S, VOLUME 15 ML. NO HARM OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MERIT CUSTOM FIT | CATHETER, INTRAVASCULAR, DIAGNOSTIC | DTL | MERIT MEDICAL SYSTEMS, INC. | H117785 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CONTRAST MEDIA |