FDA Adverse Event Malfunction Summary report: N

MERIT CUSTOM FIT

MDR report key: 2043411 · Received February 25, 2011

Report

Report Number
1721504-2011-00071
Event Type
Malfunction
Date Received
February 25, 2011
Date of Event
December 22, 2010
Report Date
February 3, 2011
Manufacturer
MERIT MEDICAL SYSTEMS, INC.
Product Code
DTL
PMA / PMN Number
K934123
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL: ONE USED SUSPECT DEVICE WAS RETURNED FOR EVAL. UPON VISUAL INSPECTION THE COMPLAINT WAS CONFIRMED. THE ROTATOR ON THE STOPCOCK WAS SEPARATED AT THE BOND BETWEEN THE COLLAR AND THE HOUSING. A REVIEW OF THE DEVICE HISTORY RECORD DID NOT REVEAL ANY EXCEPTION DOCUMENTS. THE COMPLAINT DATA BASE WAS REVIEWED AND FOUND NO SIMILAR COMPLAINTS FOR THIS LOT NUMBER. THE ROOT CAUSES OF THE FAILURE ARE ATTRIBUTED TO THE MFG BONDING PROCESS. CORRECTIVE ACTIONS HAVE BEEN IMPLEMENTED TO ADDRESS THIS TYPE OF FAILURE. THIS SUBASSEMBLY WAS BUILT PRIOR TO IMPLEMENTATION OF THE NOTED CORRECTIVE ACTIONS. EVAL METHOD: DEVICE HISTORY RECORD WAS REVIEWED, COMPLAINT DATA BASE WAS REVIEWED.

Description of Event or Problem · 1

THE ROTATOR BROKE DURING USE. INJECTOR SETTINGS: FLOW 10 ML/S, VOLUME 15 ML. NO HARM OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MERIT CUSTOM FIT CATHETER, INTRAVASCULAR, DIAGNOSTIC DTL MERIT MEDICAL SYSTEMS, INC. H117785

Patients

Seq Age Sex Outcome Treatment
1 CONTRAST MEDIA