FDA Adverse Event Malfunction Summary report: N

6800

MDR report key: 2043401 · Received March 28, 2011

Report

Report Number
1720753-2011-02888
Event Type
Malfunction
Date Received
March 28, 2011
Date of Event
March 7, 2011
Report Date
March 28, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE MFR'S SVC REP PERFORMED AN ON-SITE INVESTIGATION. THE COLLIMATOR WAS REPLACED AND ALIGNED. THE SYS WAS TESTED AND IS OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 6800 SYS HAD POOR IMAGE QUALITY. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 6800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 6800

Patients

Seq Age Sex Outcome Treatment
1