FDA Adverse Event
Malfunction
Summary report: N
ADVANCE
MDR report key: 2043398
·
Received March 15, 2011
Report
- Report Number
- 1824206-2011-01563
- Event Type
- Malfunction
- Date Received
- March 15, 2011
- Date of Event
- February 23, 2011
- Report Date
- February 23, 2011
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TECHNICIAN FOUND BRAKE BLOCK ASSEMBLY WAS MISSING BOLTS AND WORN. REPLACED BRAKE BLOCK ASSEMBLY TO RESOLVE THIS ISSUE.
Description of Event or Problem · 1
COMPLAINT ALLEGED THAT THE BRAKES WOULD NOT LOCK AND HOLD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADVANCE | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | P1145 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |