FDA Adverse Event Malfunction Summary report: N

VERSACARE

MDR report key: 2043366 · Received March 15, 2011

Report

Report Number
1824206-2011-01559
Event Type
Malfunction
Date Received
March 15, 2011
Date of Event
February 23, 2011
Report Date
February 23, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN FOUND BED IN ENGINEERING DEPARTMENT. FOUND: NURSE CALL NOT COMMUNICATING TO NURSE'S STATION DUE TO DAMAGED PIN ON SIDECOM BOARD WHERE COMMUNICATION CABLE CONNECTS. REMOVED AND REPLACED DAMAGED SIDECOM BOARD AND INSTALLED BREAKAWAY CABLE TO RESOLVE THIS ISSUE.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT NURSE CALL WAS NOT COMMUNICATING TO NURSE'S STATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VERSACARE AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. P3200

Patients

Seq Age Sex Outcome Treatment
1