ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2011-01569
- Event Type
- Malfunction
- Date Received
- March 15, 2011
- Date of Event
- February 6, 2011
- Report Date
- February 13, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. IT WAS NOTED THAT ALL KEYPAD BUTTON SYMBOLS WERE WORN. THE DOWN ARROW AND OK KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE DURING TESTING. ALL KEYPAD BUTTONS WERE OBSERVED TO SPRING BACK AS EXPECTED WHEN PRESSED. THERE WAS EVIDENCE OF CONTAMINATION FOUND UNDER ALL KEY CONTACTS.
IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. A FAMILY MEMBER REPORTED THAT THE DOWN ARROW KEYPAD BUTTON HAS BEEN SLOW TO RESPOND AND REQUIRES EXCESSIVE FORCE TO OBTAIN A RESPONSE FOR THE PAST WEEK. SHE NOTED THAT ALL OTHER KEYPAD BUTTONS ARE RESPONDING AS USUAL. THE FAMILY MEMBER REPORTED THAT THE KEYPAD BUTTONS SPRING BACK AS EXPECTED WHEN PRESSED. SHE DENIED PHYSICAL DAMAGE TO THE RUBBER KEYPAD; DENIED THAT THE PUMP HAS BEEN EXPOSED TO WATER OR CLEANING PRODUCTS; AND STATED THAT THE PATIENT WEARS THE PUMP IN VARIOUS PLACES WITH A CLIP OR CASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 13 YR |