FDA Adverse Event Malfunction Summary report: N

MICRO SAGITTAL SAW

MDR report key: 2043324 · Received March 15, 2011

Report

Report Number
1811755-2011-00814
Event Type
Malfunction
Date Received
March 15, 2011
Date of Event
February 10, 2011
Report Date
February 14, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HBE
PMA / PMN Number
K032303
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

DURING QUALITY INVESTIGATION, THE CUSTOMER'S COMPLAINT WAS CONFIRMED; THE DEVICE OVERHEATED DURING TESTING. FURTHER INVESTIGATION REVEALED CORROSION DAMAGE TO THE MOTOR AND A BROKEN BALL BEARING AND OTHER COMPONENT PARTS. THE CULPRIT PARTS WERE REPLACED AND THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

A STRYKER MICRO SAGITTAL SAW WAS SENT IN FOR REPAIR DUE TO OVERHEATING DURING TESTING. THERE WAS NO PATIENT INVOLVEMENT, NO ADVERSE EVENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICRO SAGITTAL SAW BONE CUTTING INSTRUMENT AND ACCESSORIES HBE STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK