FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2043305 · Received March 15, 2011

Report

Report Number
2531779-2011-01578
Event Type
Malfunction
Date Received
March 15, 2011
Report Date
February 15, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. DURING INVESTIGATION, IT WAS FOUND THAT THE KEYPAD BUTTONS REQUIRED EXCESSIVE FORCE TO ACTIVATE A RESPONSE. DURING INVESTIGATION, IT WAS ALSO OBSERVED THAT THE UP ARROW, DOWN ARROW AND CONTRAST BUTTONS WERE MISALIGNED. DURING INVESTIGATION IS WAS OBSERVED THAT THE BUTTONS DID NOT ALWAYS SPRING BACK WHEN PRESSED. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTRACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD IS INTERMITTENTLY RESPONDING. IT WAS ALSO REPORTED THAT THERE IS NO TEAR OR PEELING OF THE KEYPAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1