FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2043303 · Received March 15, 2011

Report

Report Number
2531779-2011-01575
Event Type
Malfunction
Date Received
March 15, 2011
Report Date
February 14, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. THE UP ARROW, OK AND CONTRAST KEYPAD BUTTONS DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE DOWN ARROW BUTTON REQUIRED EXCESSIVE FORCE TO ACTIVATE A RESPONSE. DURING INVESTIGATION, THE UP AND DOWN ARROW BUTTONS WERE OBSERVED TO BE MISALIGNED. IT WAS ALSO OBSERVED THAT ALL BUTTONS INTERMITTENTLY SPRING BACK WHEN PRESSED. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER ALL THE BUTTON CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD IS LESS RESPONSIVE. IT WAS REPORTED THAT THE PUMP WAS EXPOSED TO WATER ONE MONTH AGO (DROPPED IN TOILET).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1