FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 2043303
·
Received March 15, 2011
Report
- Report Number
- 2531779-2011-01575
- Event Type
- Malfunction
- Date Received
- March 15, 2011
- Report Date
- February 14, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. THE UP ARROW, OK AND CONTRAST KEYPAD BUTTONS DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE DOWN ARROW BUTTON REQUIRED EXCESSIVE FORCE TO ACTIVATE A RESPONSE. DURING INVESTIGATION, THE UP AND DOWN ARROW BUTTONS WERE OBSERVED TO BE MISALIGNED. IT WAS ALSO OBSERVED THAT ALL BUTTONS INTERMITTENTLY SPRING BACK WHEN PRESSED. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER ALL THE BUTTON CONTACTS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE KEYPAD IS LESS RESPONSIVE. IT WAS REPORTED THAT THE PUMP WAS EXPOSED TO WATER ONE MONTH AGO (DROPPED IN TOILET).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ONE TOUCH PING | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |