FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2043299 · Received March 15, 2011

Report

Report Number
2531779-2011-01574
Event Type
Malfunction
Date Received
March 15, 2011
Report Date
February 15, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. THE DOWN ARROW AND OK BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE UP ARROW AND CONTRAST BUTTONS REQUIRED EXCESSIVE FORCE TO ACTIVATE A RESPONSE. DURING INVESTIGATION, ALL BUTTONS WERE OBSERVED TO BE MISALIGNED. IT WAS ALSO OBSERVED THAT ALL BUTTONS DO NOT ALWAYS SPRING BACK WHEN PRESSED. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE UP ARROW BUTTON IS NOT RESPONDING. IT WAS ALSO REPORTED THERE ARE NO SIGNS OF STRUCTURAL DAMAGE AND THERE IS NO EXPOSURE TO WATER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1