FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2043286 · Received March 15, 2011

Report

Report Number
2531779-2011-01577
Event Type
Malfunction
Date Received
March 15, 2011
Report Date
February 15, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. DURING INVESTIGATION, IT WAS FOUND THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. DURING INVESTIGATION, IT WAS ALSO OBSERVED THAT THE KEYPAD BUTTONS DO NOT ALWAYS SPRING BACK. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE UP AND DOWN ARROW BUTTONS DO NOT RESPOND APPROPRIATELY. IT WAS REPORTED THAT PUMP WAS NOT EXPOSED TO MOISTURE AND THE KEYPAD IS INTACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1