FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM

MDR report key: 2043256 · Received April 7, 2011

Report

Report Number
3005075853-2011-01394
Event Type
Malfunction
Date Received
April 7, 2011
Date of Event
February 8, 2011
Report Date
February 10, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EMPTY. THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON FIRING OF THE DEVICE, THE TIP OF THE ADVANCER SLID TOWARD TISSUE STOP DURING TWO NON-CONSECUTIVE FIRING SEQUENCES CAUSING THAT THE JAWS REMAINED IN THE CLOSED. THE TRIGGER WAS MANUALLY ASSISTED IN ORDER TO OPEN THE JAWS WITHOUT ANY DIFFICULTIES NOTED; PEAR SHAPED CLIPS WERE RELEASED. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED BUT THE ORANGE INDICATOR WAS NOT VISIBLE. THIS FINDING IS NOT RELATED WITH THE INCIDENT REPORTED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC ESOPHAGECTOMY PROCEDURE, JAMMING OCCURRED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGAMAX-5MM CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA G4RN0X

Patients

Seq Age Sex Outcome Treatment
1